![]() |
Week technology ltd
|
Price: | 0.1 USD |
Payment Terms: | T/T |
Place of Origin: | Guangdong, China (Mainland) |
![]() |
|
Add to Basket | Add to My Favorites |
HiSupplier Escrow |
FOP-M260 Medical Pressure Sensor
APPLICATION
Cardiovascular: LV pressure, arterial BP
Pharmacology: Drug and Fluid injection
Neurosciences: Intracranial pressure
Spine - Intradiscal pressure
Bone - Intramedullary pressure
Urology - Bladder/Ureter pressure
MRI RFI Gating - Arterial blood pressure or LV pres-sure for image gating
Respiratory / Pulmonology
Otorhinolaryngology: Inner ear pressure
Ophthalmology: Intraocular pressure
Gastro intestinal
SPECIFICATIONS
Pressure Range 1 | -300mmHg to 300mmHg |
Resolution 2 | 0.1mmHg |
System Accuracy 3 | ±3mmHg |
Zero thermal effect 4 | <0.3mmHg/°C |
Proof pressure | >4500mmHg |
Cable Sheathing | Nylon Sheathing, OD: 0.9 mm |
Tip Termination | Bare / Sheathed / Sheathed with gel / Custom design |
Standard Sensor Length | 2 Meters |
Connector | SCAI, SCAI is a SCA connector with smart chip communi-cating calibrating data to the signal reading module |
1. Relative to atmospheric pressure
2. With a FPI-HS reading module and a filter of 30Hz applied.
3. M260 sensor and FPI-HR module : includes reproducibility (sensor/module exchange), repeatability and hysteresis, non-linearity, scale error, offset error, conditioner temperature compensation error.
4. Determined between 10°C and 70°C at atmospheric pressure (∼760mmHg)